How to use EC STER:
EC STER allows you to sterilize and simultaneously clean dirty instruments with a single treatment through an immersion procedure in its solution for the release of free radicals with reducing activity.
How to prepare EC STER
1. Pour the contents of 1 single-dose bottle into 2 liters of water and start stirring at the same time as decanting. Mains water is sufficient to
room temperature.
2. Continue stirring until the powder is completely dissolved.
3. Wait about 10 minutes. The solution will have the characteristic pale yellow color that confirms the beginning of sterilizing activity. (If the tap water has a degree of
hardness greater than 20° F the solution may appear cloudy).
4. Immerse the dirty instruments in the solution and leave them immersed for about 10 minutes. Make sure that the solution penetrates all cavities.
After 2 minutes of immersion, all surfaces that have come into contact with the solution are sterile. Cleansing is complete after about 10 minutes (depending on the degree of soiling).
5. Remove instruments, rinse thoroughly and dry with sterile technique.
N.B.
- For the preparation of the active solution, use clean containers, which do not contain traces of surfactant detergents or acidic substances.
- When handling instruments with tungsten components, it is recommended to
Respect the indicated contact times.
- Keep the basin containing the solution covered to avoid prolonged contact with oxygen in the air.